<?xml version="1.0" encoding="windows-1252"?><SEC xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:noNamespaceSchemaLocation="http://si.ksc.nasa.gov/sidownloads/xml/specsintactSEC.xsd"><MTA NAME="SUBFORMAT" CONTENT="NEW"/><HDR><AST/><BRK/>
USACE / NAVFAC / AFCESA / NASA      UFGS-11 74 00 (April 2006)<BRK/>
                                    ------------------------------<BRK/>
Preparing Activity:  <PRA>NAVFAC</PRA>         Replacing without change<BRK/>
                                    UFGS-11744 (August 2004)<BRK/>
<BRK/>
<HL4>UNIFIED FACILITIES GUIDE SPECIFICATIONS</HL4><BRK/>
<BRK/>
<HL4>References are in agreement with UMRL dated January 2009</HL4><BRK/>
<AST/><BRK/></HDR>
<BRK/>
<SCN>SECTION 11 74 00</SCN><BRK/>
<BRK/>
<STL>DENTAL EQUIPMENT</STL><BRK/>
<DTE>04/06</DTE><BRK/>
<NTE><BRK/>
<AST/><BRK/>
<NPR>NOTE:  This guide specification covers the requirements for <SCP>dental equipment 
including dental operating lights and similar related dental equipment</SCP>.<BRK/>
<BRK/>
Edit this guide specification for project specific requirements by adding, deleting, 
or revising text.  For bracketed items, choose applicable items(s) or insert 
appropriate information.<BRK/>
<BRK/>
Remove information and requirements not required in respective project, whether 
or not brackets are present.<BRK/>
<BRK/>
Comments and suggestions on this guide specification are welcome and should 
be directed to the technical proponent of the specification.  A listing of <URL HREF="http://65.204.17.188/report/ufgs.html">technical 
proponents</URL>, including their organization designation and telephone number, is 
on the Internet.<BRK/>
<BRK/>
Recommended changes to a UFGS should be submitted as a  <URL HREF="http://65.204.17.188/projnet/cms/public.html">Criteria Change Request 
(CCR)</URL>.</NPR><BRK/>
<AST/><BRK/></NTE>
<NTE><BRK/>
<AST/><BRK/>
<NPR>NOTE: TO DOWNLOAD UFGS GRAPHICS<BRK/>
<BRK/>
Go to <URL HREF="http://www.wbdg.org/ccb/NAVGRAPH/graphtoc.pdf">http://www.wbdg.org/ccb/NAVGRAPH/graphtoc.pdf</URL>.</NPR><BRK/>
<AST/><BRK/></NTE>
<BRK/>
<PRT><TTL>PART 1   GENERAL</TTL><BRK/>
<BRK/>
<SPT><TTL>1.1   REFERENCES</TTL><BRK/>
<NTE><BRK/>
<AST/><BRK/>
<NPR>NOTE:  This paragraph is used to list the publications cited in the text of 
the guide specification. The publications are referred to in the text by basic 
designation only and listed in this paragraph by organization, designation, 
date, and title.<BRK/>
 <BRK/>
Use the Reference Wizard's Check Reference feature when you add a RID outside 
of the Section's Reference Article to automatically place the reference in the 
Reference Article.  Also use the Reference Wizard's Check Reference feature 
to update the issue dates.<BRK/>
 <BRK/>
References not used in the text will automatically be deleted from this section 
of the project specification when you choose to reconcile references in the 
publish print process.</NPR><BRK/>
<AST/><BRK/></NTE>
<BRK/>
<TXT>The publications listed below form a part of this specification to the extent referenced.  The publications are 
referred to within the text by the basic designation only.</TXT><BRK/>
<BRK/>
<REF><ORG>ASTM INTERNATIONAL (ASTM)</ORG><BRK/><BRK/><RID>ASTM A 167</RID><RTL>(1999; R 2004) Standard Specification for Stainless and Heat-Resisting Chromium-Nickel Steel Plate, Sheet, and Strip</RTL><BRK/><BRK/><RID>ASTM A 568/A 568M</RID><RTL>(2007a) Standard Specifications for Steel, Sheet, Carbon, and High-Strength, Low-Alloy, Hot-Rolled and Cold-Rolled, General Requirements for</RTL><BRK/><BRK/><RID>ASTM B 221</RID><RTL>(2008) Standard Specification for Aluminum and Aluminum-Alloy Extruded Bars, Rods, Wire, Profiles, and Tubes</RTL><BRK/><BRK/><RID>ASTM B 221M</RID><RTL>(2007) Standard Specification for Aluminum and Aluminum-Alloy Extruded Bars, Rods, Wire, Profiles, and Tubes (Metric)</RTL><BRK/><BRK/><RID>ASTM E 308</RID><RTL>(2006) Computing the Colors of Objects by Using the CIE System</RTL><BRK/><BRK/></REF><REF><ORG>COMPRESSED GAS ASSOCIATION (CGA)</ORG><BRK/><BRK/><RID>CGA G-10.1</RID><RTL>(2008) Commodity Specification for Nitrogen</RTL><BRK/><BRK/><RID>CGA G-7.1</RID><RTL>(2004) Commodity Specification for Air</RTL><BRK/><BRK/></REF></SPT><SPT><TTL>1.2   RELATED WORK SPECIFIED IN OTHER SECTIONS</TTL><BRK/>
<BRK/>
<TXT>Conform to the provisions of Section <SRF>11 70 00</SRF> GENERAL REQUIREMENTS FOR MEDICAL AND DENTAL EQUIPMENT.  Provide 
final utility connections and utility service to equipment under Sections <SRF>23 03 00.00 20</SRF> BASIC MECHANICAL MATERIALS 
AND METHODS, <SRF>22 00 00</SRF> PLUMBING SYSTEMS, <SRF>22 60 70</SRF> GAS AND VACUUM SYSTEMS FOR HEALTHCARE FACILITIES, 
<SRF>26 00 00.00 20</SRF> BASIC ELECTRICAL MATERIALS AND METHODS, and <SRF>26 20 00</SRF> INTERIOR DISTRIBUTION SYSTEM."</TXT><BRK/>
<BRK/></SPT>
<SPT><TTL>1.3   SUBMITTALS</TTL><BRK/>
<NTE><BRK/>
<AST/><BRK/>
<NPR>NOTE:  Review submittal description (SD) definitions in Section 01 33 00 SUBMITTAL 
PROCEDURES and edit the following list to reflect only the submittals required 
for the project.  Submittals should be kept to the minimum required for adequate 
quality control.<BRK/>
<BRK/>
A “G” following a submittal item indicates that the submittal requires Government 
approval.  Some submittals are already marked with a “G”.  Only delete an existing 
“G” if the submittal item is not complex and can be reviewed through the Contractor’s 
Quality Control system.  Only add a “G” if the submittal is sufficiently important 
or complex in context of the project.<BRK/>
<BRK/>
For submittals requiring Government approval on Army projects, a code of up 
to three characters within the submittal tags may be used following the "G" 
designation to indicate the approving authority.  Codes for Army projects using 
the Resident Management System (RMS) are:  "AE" for Architect-Engineer; "DO" 
for District Office (Engineering Division or other organization in the District 
Office); "AO" for Area Office; "RO" for Resident Office; and "PO" for Project 
Office.  Codes following the "G" typically are not used for Navy,  Air Force, 
and NASA projects.<BRK/>
<BRK/>
Choose the first bracketed item for Navy, Air Force and NASA projects, or choose 
the second bracketed item for Army projects.</NPR><BRK/>
<AST/><BRK/></NTE>
<BRK/>
<TXT>Government approval is required for submittals with a "G" designation; submittals not having a "G" designation 
are [for Contractor Quality Control approval.][for information only.  When used, a designation following the 
"G" designation identifies the office that will review the submittal for the Government.]  The following shall 
be submitted in accordance with Section <SRF>01 33 00</SRF> SUBMITTAL PROCEDURES:</TXT><BRK/>
<BRK/>
<TXT>Conform to the submittal provisions of Section <SRF>11 70 00</SRF> GENERAL REQUIREMENTS FOR MEDICAL AND DENTAL EQUIPMENT.</TXT><BRK/>
<BRK/>
<LST><SUB>SD-02 Shop Drawings</SUB></LST><BRK/>
<BRK/>
<ITM>Schedule of dental equipment <SUB>items</SUB></ITM><BRK/>
<BRK/>
<LST><SUB>SD-03 Product Data</SUB></LST><BRK/>
<BRK/>
<ITM><SUB>Dust collector</SUB> base pedestal[; <SUB>G</SUB>][; <SUB>G, [_____]</SUB>]</ITM><BRK/>
<BRK/>
<ITM><SUB>Air compressor</SUB>, dental[; <SUB>G</SUB>][; <SUB>G, [_____]</SUB>]</ITM><BRK/>
<BRK/>
<ITM><SUB>Light</SUB>, dental operating[; <SUB>G</SUB>][; <SUB>G, [_____]</SUB>]</ITM><BRK/>
<BRK/>
<ITM><SUB>Utility service center</SUB>[; <SUB>G</SUB>][; <SUB>G, [_____]</SUB>]</ITM><BRK/>
<BRK/></SPT>
</PRT><PRT><TTL>PART 2   PRODUCTS</TTL><BRK/>
<BRK/>
<SPT><TTL>2.1   MATERIALS</TTL><BRK/>
<BRK/>
<SPT><TTL>2.1.1   Stainless Steel</TTL><BRK/>
<BRK/>
<TXT><RID>ASTM A 167</RID>, Class 301, 302, or 304.  Exposed surfaces shall have a No. 4 finish.</TXT><BRK/>
<BRK/></SPT>
<SPT><TTL>2.1.2   Aluminum Alloy</TTL><BRK/>
<BRK/>
<TXT><MET><RID>ASTM B 221M</RID></MET><ENG><RID>ASTM B 221</RID></ENG>; provide alloy equivalent in ultimate tensile, yield, and shear strengths to Alloy 6063-T5 
or 6063-T6, as applicable.</TXT><BRK/>
<BRK/></SPT>
<SPT><TTL>2.1.3   Carbon Steel</TTL><BRK/>
<BRK/>
<TXT><RID>ASTM A 568/A 568M</RID>; cold-rolled, stretcher level sheets<MET> 1.5 mm thick</MET><ENG> 16 gage</ENG> and lighter with a commercial bright 
finish.</TXT><BRK/>
<BRK/></SPT>
</SPT><SPT><TTL>2.2   <SUB>ITEMS</SUB></TTL><BRK/>
<BRK/>
<SPT><TTL>2.2.1   Special Items</TTL><BRK/>
<NTE><BRK/>
<AST/><BRK/>
<NPR>NOTE:  Insert a description of the item in the first two blanks, as sufficient 
to determine the type of product required.  Avoid repetition of detailed information 
which appears in the salient features listed thereafter.</NPR><BRK/>
<AST/><BRK/></NTE>
<BRK/>
<TXT>[Item _____, _____]:  The unit shall be [_____].  Include the following:</TXT><BRK/>
<BRK/>
<ITM INDENT="-0.33">[a.  _____]</ITM><BRK/>
<BRK/></SPT>
<SPT><TTL>2.2.2   <SUB>Dust Collector</SUB>, Base Pedestal</TTL><BRK/>
<BRK/>
<TXT>Provide a cabinet-enclosed, floor-mounted type suitable for use with grinding and polishing hoods in dental laboratories.  
Include the following:</TXT><BRK/>
<BRK/>
<ITM INDENT="-0.33">a.  A capacity for at least four hoods.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">b.  A self-contained, 220-volt, 60 hertz (Hz),<MET> 750 watt</MET><ENG> one-horsepower (hp)</ENG> electric motor.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">c.  Overall dimensions:  Approximately<MET> 625 mm</MET><ENG> 25 inches</ENG> high by <MET> 1050 mm</MET><ENG> 42 inches</ENG> long by<MET> 375 
mm</MET><ENG> 15 inches</ENG> wide.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">d.  Filters:  Permanent (cleanable) type; 99 percent-high efficiency.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">e.  Cabinet design:  Removable snap-on panels for easy access to filters; equipped with a footpedal 
shaker on the outside.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">f.  Ductwork from collector unit to hoods:  Include flexible links, branch connectors, and end 
caps.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">g.  Exhaust:  Vented to the exterior of the building.</ITM><BRK/>
<BRK/></SPT>
<SPT><TTL>2.2.3   <SUB>Air Compressor</SUB>, Dental</TTL><BRK/>
<BRK/>
<TXT>Provide an air compressor capable of providing compressed air and meeting the following criteria:</TXT><BRK/>
<NTE><BRK/>
<AST/><BRK/>
<NPR>NOTE:  Specify sufficient cfm of compressed air to satisfy the flow requirements 
of the combined equipment served under operational conditions.</NPR><BRK/>
<AST/><BRK/></NTE>
<BRK/>
<ITM INDENT="-0.33">a.  Capacity:  Capable of providing [_____]<MET> liters per minute (lpm)</MET><ENG> cubic feet per minute (cfm)</ENG>
 of compressed air at a residual pressure of<MET> 650 kilopascals</MET><ENG> 80 pounds per square inch (psi)</ENG>
.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">b.  Isolation:  Mount pumps and similar vibrating and noise-producing components on the isolator.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">c.  Electrical service characteristics:  [120/208 volts, [single] [three]-phase] [240 volts, 
[single] [three]-phase], [60] [50] Hz.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">d.  Quality of air:</ITM><BRK/>
<MET><TBL><THD><BRK/>
  <HL1>CONTAMINANT</HL1>                           <HL1>LIMIT</HL1><BRK/></THD>
<BRK/>
  Water                    Dry to a pressure dewpoint of 2 degrees C<BRK/>
                           at a minimum of 720 kilopascals<BRK/>
                           (<RID>CGA G-7.1</RID>)<BRK/>
<BRK/>
  Condensed hydrocarbons   A maximum of 0.1 part per million (ppm)<BRK/>
                           by weight or 0.1 milligram per liter<BRK/>
                          (<RID>CGA G-10.1</RID>)<BRK/>
<BRK/>
  Permanent particulates   Less than 1.0 ppm by weight or 1.0<BRK/>
                           milligram per liter (<RID>CGA G-10.1</RID>)<BRK/></TBL>
</MET><ENG><TBL><THD><BRK/>
  <HL1>CONTAMINANT</HL1>                           <HL1>LIMIT</HL1><BRK/></THD>
<BRK/>
  Water                    Dry to a pressure dewpoint of 35 degrees F<BRK/>
                           at a minimum of 90 pounds per square<BRK/>
                           inch gage (psig) (<RID>CGA G-7.1</RID>)<BRK/>
<BRK/>
  Condensed hydrocarbons   A maximum of 0.1 part per million (ppm) by<BRK/>
                           weight or 0.1 milligram per liter (<RID>CGA G-10.1</RID>)<BRK/>
<BRK/>
  Permanent particulates   Less than 1.0 ppm by weight or 1 0<BRK/>
                           milligram per liter (<RID>CGA G-10.1</RID>)<BRK/></TBL>
</ENG><BRK/></SPT>
<SPT><TTL>2.2.4   <SUB>Light</SUB>, Dental Operating</TTL><BRK/>
<BRK/>
<TXT>Provide a twin lamphead (open reflector-type) suspended on a counterbalanced arm from a corrosion-resistant finish 
carriage that rides on a ceiling track.  Include the following:</TXT><BRK/>
<BRK/>
<ITM INDENT="-0.33">a.  Ceiling track:  For surface mounting to a finished ceiling.  The track shall be approximately<MET>
 1980 mm</MET><ENG> 78 inches</ENG> long.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">b.  Carriage:  Track-mounted; to travel silently and smoothly along ceiling track length without 
binding and to stop without drifting.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">c.  Light bulb:  Quartz-halogen; 12 volt; rated to exceed a 1,800 hour lifespan.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">d.  On/off switch:  Yoke-mounted; elongated bar, toggle-type.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">e.  Lighthead:  The lighthead shall be light in weight and shall be easily positioned.  The 
lighthead shall feature a multifaceted reflector or other suitable means of simulating a multipoint 
light source to minimize the effect of blocking shadows.  The lighthead shall be adjustable 
to illuminate all areas of the oral cavity. The lighthead shall be capable of being positioned 
at all points within the base of a spherical segment whose radius is equal to a 90 degree solid 
angle located at the imaginary center of the oral cavity and whose axis is coincident with a 
sagittal axis through the imaginary center with the patient reclined in a dental chair (chair 
back positioned at 45 degrees).</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">f.  Illumination intensity:  The light shall feature a minimum of three intensity settings.  
The lowest intensity shall be<MET> 8600 to 12,900 lux</MET><ENG> 800 to 1,200 footcandles</ENG>, and the highest intensity 
shall be a minimum of<MET> 21,500 lux</MET><ENG> 2,000 footcandles</ENG> at the specified optimum focal length between<MET>
 625 and 800 mm</MET><ENG> 25 and 32 inches</ENG> from the oral cavity.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">g.  Color properties:  The light shall be capable of providing a color-corrected natural white 
light suitable for critical shade discrimination characterized by a correlated color temperature 
in the range from 3,800 to 5,500 degrees Kelvin (K) with a broad spectrum output containing 
all visible wavelengths.  The emitted light shall have <RID>ASTM E 308</RID> color within an area described 
by a rectangle on the CIE diagram with the range of coordinates describing the rectangle defined 
by the following U and V prime (') values.</ITM><BRK/>
<TBL><THD><BRK/>
                       <HL1> U'</HL1>             <HL1>V'</HL1><BRK/></THD>
<BRK/>
                      0.2350         0.4975<BRK/>
                      0.2250         0.5225<BRK/>
                      0.1998         0.4925<BRK/>
                      0.2158         0.4700<BRK/></TBL>
<BRK/>
<ITM INDENT="-0.33">h.  Light bulb replacement:  The light bulb shall be field replaceable without tools or with 
a tool supplied by the manufacturer of the light.  If a tool is supplied by the manufacturer 
of the light, there shall be a means of attaching the tool to the control arm of the light.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">i.  Electrical characteristics:  120 volts; single-phase; 60 Hz.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">j.  Enameled components:  Finish enameled components in a light manufacturer's standard color.</ITM><BRK/>
<BRK/>
<ITM INDENT="-0.33">k.  Mounting hardware:  Include all mounting hardware to ensure that the track and light are 
securely mounted.</ITM><BRK/>
<BRK/></SPT>
<SPT><TTL>2.2.5   <SUB>Utility Service Center</SUB></TTL><BRK/>
<NTE><BRK/>
<AST/><BRK/>
<NPR>NOTE:  Utility Service Center shall be used in conjunction with Item D8255, 
Dental Delivery System, a Government-furnished and -installed item. Transfer 
the template to the project drawings; in doing so, delete the words "Sketch 
No. 11744-1," but retain the title of the template.</NPR><BRK/>
<AST/><BRK/></NTE>
<BRK/>
<TXT>Install utilities indicated on "Dental Delivery System Utility Connection Template D7090 - Utility, Center Dental."  
Provide manual shutoff valves, one each for the air and water supply.  Both valves shall be compatible with the 
specified inlet pipes, and both valves shall have a<MET> 20 mm</MET><ENG> 3/8 inch</ENG> compression fitting on the outlet.</TXT><BRK/>
<BRK/></SPT>
</SPT></PRT><PRT><TTL>PART 3   EXECUTION</TTL><BRK/>
<BRK/>
<SPT><TTL>3.1   INSTALLATION</TTL><BRK/>
<BRK/>
<TXT>Conform to the installation provisions of Section <SRF>11 70 00</SRF> GENERAL REQUIREMENTS FOR MEDICAL AND DENTAL EQUIPMENT, 
for the items specified herein.  Install the items specified herein at the locations indicated on the drawings.</TXT><BRK/>
<BRK/></SPT>
<SPT><TTL>3.2   FIELD QUALITY CONTROL</TTL><BRK/>
<BRK/>
<SPT><TTL>3.2.1   Inspection</TTL><BRK/>
<BRK/>
<TXT><TST>Examine each item for visual defects and conformance to specifications.</TST></TXT><BRK/>
<BRK/></SPT>
<SPT><TTL>3.2.2   Tests</TTL><BRK/>
<BRK/>
<TXT><TST>Test each item to ensure that the equipment is operational and conforms to specifications.</TST></TXT><BRK/>
<BRK/></SPT>
</SPT></PRT>    <END/><BRK/></SEC>